Water Ph TDS Testing Devices

Description

Water Ph TDS Testing Devices

Details of Ph Testing Device

The HI98107 features a large multi-level LCD which displays both pH and temperature readings simultaneously. The pH readings are displayed with a 0.1 pH resolution and with an accuracy of ±0.1 pH. The pH range of the HI98107 is from 0.0 to 14.0 pH. An exposed temperature sensor allows for rapid automatic temperature compensated pH measurements. Temperature can be set to display in ℃ or ℉. The LCD screen has stability and calibration tag indicators. The battery percent level is displayed at start up alerting the user to the remaining battery power that is available.

The pH electrode uses a unique reference cell design that utilizes an extractable cloth junction. A clogged junction will typically exhibit erratic readings that will not stabilize. Simply extracting a small portion of cloth from the water pH teter will clear any clog by exposing a fresh, clean surface. The ability to clear any clogging will extend the life of the pH electrode.

Calibration is performed by pressing the dedicated calibration button. Once in calibration mode the pH electrode can be calibrated to 1 or 2 points using the pre-programmed standard buffers (pH 4.01, 7.01,10.01). When the standard value is recognized and stability is reached, the meter automatically accepts and stores the calibration value.

Setup options include automatic shut off timing and temperature unit displayed. Automatic shut-off options are to turn the water pH meter off after 8 minutes, 60 minutes, or disabled. To enter the setup mode the user presses a button located in the battery compartment.

Details TDS Testing Device

The DiST 1 features a large multi-level LCD which displays both TDS and temperature readings simultaneously. The TDS readings are displayed with a 1 ppm resolution and with an accuracy of ±2% full scale. The TDS range of the HI98301 is from 0 to 2000 ppm. An exposed temperature sensor allows for rapid automatic temperature compensated TDS measurements. Temperature can be set to display in ℃ or ℉. The LCD screen has stability and calibration tag indicators. The battery percent level is displayed at start up alerting the user to the remaining battery power that is available.

TDS readings are performed by an amperometric graphite electrode that resists oxidation, providing better repeatability in measurements. The use of graphite also reduces the polarization effect common with amperometric testers that use two stainless steel pins.

Calibration is performed by pressing the dedicated calibration button. Once in calibration mode submerge the probe in 1382 ppm calibration standard. When the standard value is recognized and stability reached, the meter will automatically accept and store the calibration value.

Setup options include automatic shut-off timing and temperature unit displayed. Automatic shut-off options are to turn the meter off after 8 minutes, 60 minutes, or disabled. To enter the setup mode the user presses a button located in the battery compartment.